飞利浦BV Libra C臂故障信息查询

飞利浦BV Libra C臂故障信息查询

error
  mnemonic:

displayed STUIF texts
  (dismsg.c):

LOG file text and remarks (co_error.c):

019

X-ray generator not available
  Switch system off/on. Call
  service if problem is
  persistent

Invalid error id or no error message   defined !!!

020

Warning: system parameters
  are reset. Call service

All DATABASE data has been reset or new   DATABASE has
  been installed. Calibrated BV-scope data has been removed and
  replaced with the default settings. Perform the following actions:
  Perform Program Automatic install configuration and
  examination. Check the program manual functions. Perform the
  adjustments X-ray tube: Filament and Dose output. Perform the
  adjustments Image processor. Perform the adjustments
  Collimator (excluding laser alignment). Perform the adjustment
  II/TV: Electrical-Dose rate.

021

Warning: system parameters
  are reset. Call service

Configuration data has been reset. The   default settings are now
  used. Perform the following action: Using the Program Automatic
  function, install Configuration.

022

Warning: system parameters
  are reset. Call service

No BV-scope calibration has been   performed, or calibration data
  has been reset. The default settings are now used. Perform the
  following actions: Perform the adjustments X-ray tube: Dose
  output. Perform the adjustments Image processor. Perform
  adjustments Collimator (excluding laser alignment). Perform the
  adjustment II/TV: Electrical-Dose rate.

023

Warning: system parameters
  are reset. Call service

The APF data has been reset. The default   settings are now used.
  Perform the following action: Using the Program Automatic
  function, install Examination. Check program manual functions:
  Examination names and APF settings.

024

Warning: system parameters
  are reset. Call service

The heat integrator data has been reset.   This will correct itself
  automatically within 5 minutes - no further action required.

025

Warning: system parameters
  are reset. Call service

The current patient dose data has been   reset. No action required.

026

Warning: system parameters
  are reset. Call service

Annabelle DATABASE data has been reset. The   default settings
  are now used. No action required.

027

Warning: system parameters
  are reset. Call service

The dose output calibration data has been   reset. Perform the
  adjustment for the X-ray tube: Dose output.

028

Warning: system parameters
  are reset. Call service

The tank history data has been reset.   NOTE: The old tank history
  data has been deleted. No action required.

029

Warning: system parameters
  are reset. Call service

The energy storage history data has been   reset. NOTE: The old
  energy storage history data has been deleted. No action required.

030

Warning: system parameters
  are reset. Call service

NOTE: The dose report of the current   acquisition patient is
  suspect. No action required.

031

Warning: system parameters
  are reset. Call service

NOTE: The filament adjustment data has   been reset. Perform the
  filament adjustment procedure.

<none>

Mobile Viewing Station not
  available. Switch system
  off/on. Call service if problem
  is persistent

<none>

<none>

Mobile Viewing Station in
  service mode. Switch off/on
  to enter normal mode

<none>

050

System initialization problem
  Switch system off/on. Call
  service if problem is
  persistent

A system startup problem occurred. Switch   the system off and
  on. If the problem persists, check if the Mobile Viewing Station
  and the Stand are correctly connected. Check the Stand-Mobile
  Viewing Station cable.

error
  mnemonic:

displayed   STUIF texts
  (dismsg.c):

LOG   file text and remarks (co_error.c):

051

Software   version mismatch
  Call service

The   software version installed on the stand is not compatible with
  the version installed on the mobile viewing station. Install the
  latest version of the software on either the stand or the DFI.

052

System   not available
  Switch system off/on
  Call service if problem is
  persistent

The   video frequency of the stand is not compatible with the video
  frequency of the mobile viewing station. Call PMS helpdesk.

053

Subsystem   error, system still
  available. It's possible to
  continue

The   DFI is not responding within the allowed time, causing a
  timeout. Switch the system off and on. Check the DFI unit for
  errors.

054

APF   parameters not
  accepted. The settings for
  current selection are invalid
  Call service

The   APF data is invalid due to an incompatibility between the
  APF data and the IP set. Perform Program Automatic Install
  Examination.

055

APF   parameters not
  accepted. The settings for
  current selection are invalid
  Call service

Unknown   IP set. Check the application data for errors. Load DFI
  IP sets.

<none>

Storage   speed not accepted

Storage   speed not selectable
  during X-ray

The   selected frame speed is not acceptable during X-ray

056

Subsystem   error, system still
  available. It's possible to
  continue

The   collimator control board is not responding within the allowed
  time, causing a timeout. Check the collimator device, board and
  cabling. If the problems persist, replace the SHA5 collimator
  board.

057

Subsystem   error, system still
  available. It's possible to
  continue

The   generator control board is not responding within the allowed
  time, causing a timeout. Check the generator device, board and
  cabling. Check if the Stand software is operational. If the
  problems persist, replace the SHA4 generator board.

058

Subsystem   error, system still
  available. It's possible to
  continue

The   TVI sub-system is not responding within the allowed time,
  causing a timeout. Check the II, the camera, pulse master board
  and cabling. Check if the Stand software is operational. If the
  problems persist, replace the pulse master board (SHA3).

059

Subsystem   error, system still
  available. It's possible to
  continue

The   pulse master control is not responding within the allowed
  time, causing a timeout. Check the board, the cabling and Stand
  software. If the problems persist, replace the board (SHA3).

060

HIP   time exceeded, low dose
  fluoro activated.

Maximum   hip time reached so hip switched off.

061

Warning:   wrong Viewing
  Station connected.
  Connect the matching
  Viewing Station

Incorrect   DFI connected with the SUCO. Find the correct DFI and
  connect this with the SUCO. (If DFI software Rev is 1.8.1 or less
  implement FCO71800067. This FCO is only applicable for BV
  Libra R1.2).

062

Image   synchronization
  problem. Call service if
  problem is persistent

No   sync detected between XTV and DFI. Check XTV camera and
  the stand-mobile viewing station cable.

<none>

Dose   limitation active! only
  low dose fluoro. Reduced
  performance.

Dose   limitation active, only low dose fluoro. Reduced
  performance.

<none>

Dose   limitation active! kV
  limited. Reduced
  performance.

Dose   limitation active, kV limited. Reduced performance.

<none>

Press   H/F switch!

<none>

<none>

Keep   H/F switch pressed ...

<none>

<none>

Adjustment   in progress ...

<none>